Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman DM10XL: (1 year) CAL VER /5500 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 21 NUMBER OF LINES: 107 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # dm10xlv.bmp # dm10xllc.bmp # dm10xlhc.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: None 1.004 DISP Temperature: 0 - 50degC 1.004 DISP Humidity: <70% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 PIC dm10xlv 1.007 HEAD {AC VOLTAGE} 1.008 HEAD AC VOLTAGE: {200V Range} 1.009 5500 200 190.0V 1% 0.4U 60H SI 2W 2.001 HEAD AC VOLTAGE: {750V Range} 2.002 5500 750 700V 1.5% 4U 60H SI 2W 3.001 HEAD {DC VOLTAGE} 3.002 HEAD DC VOLTAGE: {1000V Range} 3.003 5500 1000 1000V 0.7% 1U 2W 4.001 HEAD DC VOLTAGE: {200V Range} 4.002 5500 200 190.0V 0.7% 0.05/ 2W 5.001 HEAD DC VOLTAGE: {20V Range} 5.002 5500 20 19.00V 0.7% 0.05/ 2W 6.001 HEAD DC VOLTAGE: {2V Range} 6.002 5500 2 1.900V 0.7% 0.05/ 2W 7.001 HEAD DC VOLTAGE: {200mV Range} 7.002 5500 200 190.0mV 0.7% 0.05/ 2W 8.001 HEAD {RESISTANCE} 8.002 HEAD RESISTANCE: {20 MOhm Range} 8.003 5500 20 19.00MZ 3% 0.01U 2W 9.001 HEAD RESISTANCE: {2 MOhm Range} 9.002 5500 2 1.900MZ 1% 0.001U 2W 10.001 HEAD RESISTANCE: {200 kOhm Range} 10.002 5500 200 190.0kZ 1% 0.1U 2W 11.001 HEAD RESISTANCE: {20 kOhm Range} 11.002 DISP Connect the 5500A to the UUT as follows: 11.002 DISP [32] 5500A AUX HI to UUT V/OHM 11.002 DISP [32] 5500A AUX LO to UUT COM 11.002 DISP [32] 5500A NORMAL HI to UUT V/OHM 11.002 DISP [32] 5500A NORMAL LO to UUT COM 11.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 11.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 11.003 5500 20 19.00kZ 1% 0.01U CW 12.001 HEAD RESISTANCE: {2 kOhm Range} 12.002 5500 2 1.900kZ 1% 0.001U CW 13.001 HEAD RESISTANCE: {200 Ohm Range} 13.002 5500 200 190.0Z 1% 0.3U CW 14.001 HEAD {DIODE TESTER} 14.002 PIC dm10xlv 14.003 DISP Set the DMM FUNCTION/RANGE switch to DIODE. 14.004 5500 0.5V S 2W 14.005 EVAL Does {the display read approx. 0.500 (diode junction)}? 15.001 HEAD {DC CURRENT} 15.002 PIC dm10xllc 15.003 HEAD DC CURRENT: {200uA Range} 15.004 5500 200 190.0uA 1% 0.1U 2W 16.001 HEAD DC CURRENT: {2mA Range} 16.002 5500 2 1.900mA 1% 0.1U 2W 17.001 HEAD DC CURRENT: {20mA Range} 17.002 5500 20 19.00mA 1% 0.1U 2W 18.001 HEAD DC CURRENT: {200mA Range} 18.002 5500 200 190.0mA 1% 0.1U 2W 19.001 PIC dm10xlhc 19.002 HEAD DC CURRENT: {10A Range} 19.003 5500 10 9.90A 2% 0.03U 2W 20.001 HEAD {SAFETY TESTER} 20.002 DISP Set the DMM FUNCTION/RANGE switch to SAFETY TESTER. 20.002 DISP 20.002 DISP *NOTE* this function does not operate off the internal 20.002 DISP [32] DMM battery, instead it is powered from the 20.002 DISP [32] voltage source being measured. The LCD display 20.002 DISP [32] will be BLANK. 20.003 PIC dm10xlv 20.004 5500 25V S 2W 20.005 EVAL Do {the LEDs indicate a voltage between 24 - 50}? 21.001 END